Profibus Communications
95
Instruction Manual
Section 12: Profibus Communications
LIQ-MAN-56
April 2017
Section 12.0 Profibus Communications
12.1
General
The 56 can communicate using Profibus DP-V1 by purchasing a 56 with a Profibus
communication board installed (model code 56-PC-02-DP), or by adding a Profibus
communications board to the standard 56 with HART communications.
The standard HART 56 analyzer can be converted to Profibus DP by adding a Profibus board.
There are two different Profibus boards for the 56, which must be chosen to be compatible with
the main board software version of the 56.
•
If the 56 software version begins with a 2 (2.XX), Profibus board PN 24285-02 should be chosen.
•
If the 56 software version begins with a 1 (1.XX), Profibus board PN 24285-01 should be chosen.
A 56 with Profibus communications retains all the analog and discrete functionality of the
standard HART 56 transmitter. The four 4 to 20 mA analog outputs can be configured as analog
outputs or PID control outputs.
The four discrete relay outputs can be configured for concentration control (high/low alarms
with or without a delay timer), event based activation, timer based activation, or TPC (Duty
Cycle/ Pulse Width Modulation) control.
12.2
Profibus Features
As outlined in the Device Description Data Field (GSD), the 56 has the following Profibus DP-
V1 identification and capabilities:
Identification:
GSD Revision: 5
Vendor Name: Rosemount
Model Name: MODEL 56 PROFIBUS DP
Revision: V1.00
Identification Number: 0x10AA
Hardware Release: V1.00
Software Release: V1.00
Implementation Type: VPC3+C
Device: DP-V1 Slave
Baud Rates Supported:
All Profibus DP baud rates are supported, from 9.6 kbps to 12 Mbps.
Additional Capabilities:
• The 56 Profibus board supports the
automatic transmission rate recognition.
It
automatically sets its baud rate to the transmission rate of the Profibus network,
specified by the master.
Содержание Rosemount 56
Страница 1: ...Instruction Manual LIQ MAN 56 Rev D April 2017 Rosemount 56 Advanced Dual Input Analyzer ...
Страница 2: ......
Страница 10: ...iv Table of Contents Instruction Manual April 2017 LIQ MAN 56 ...
Страница 30: ...Section 2 Installation Instruction Manual April 2017 LIQ MAN 56 20 Installation ...
Страница 40: ...30 Wiring Section 3 Wiring Instruction Manual April 2017 LIQ MAN 56 ...
Страница 64: ...Section 6 Programming Measurements Instruction Manual April 2017 LIQ MAN 56 54 ...
Страница 76: ...Section 8 Time Proportional Control Instruction Manual April 2017 LIQ MAN 56 66 ...
Страница 124: ...Section 13 Maintenance Instruction Manual April 2017 LIQ MAN 56 114 ...
Страница 126: ...Section 14 Return of Material Instruction Manual April 2017 LIQ MAN 56 116 ...