Example: Cutoff interaction with AO Cutoff higher than Gas Standard Volume Flow
Cutoff
Configuration:
•
mA Output Process Variable
for the primary mA output:
Gas Standard Volume Flow Rate
•
Frequency Output Process Variable
:
Gas Standard Volume Flow Rate
•
AO Cutoff
for the primary mA output:
15
SLPM (standard liters per minute)
•
Gas Standard Volume Flow Cutoff
:
10
SLPM
Result:
•
If the gas standard volume flow rate drops below 15 SLPM but not below 10 SLPM:
-
The primary mA output will report zero flow.
-
The frequency output will report the actual flow rate, and the actual flow rate will
be used in all internal processing.
•
If the gas standard volume flow rate drops below 10 SLPM, both outputs will report
zero flow, and 0 will be used in all internal processing.
4.4
Configure Flow Direction
Display
Not available
ProLink II
ProLink
>
Configuration
>
Flow
>
Flow Direction
ProLink III
Device Tools
>
Configuration
>
Process Measurement
>
Flow
Field Communicator
Configure
>
Manual Setup
>
Measurements
>
Flow
>
Flow Direction
Overview
Flow Direction
controls how forward flow and reverse flow affect flow measurement and
reporting.
Flow Direction
is defined with respect to the flow arrow on the sensor:
•
Forward flow (positive flow) moves in the direction of the flow arrow on the sensor.
•
Reverse flow (negative flow) moves in the direction opposite to the flow arrow on
the sensor.
Tip
Micro Motion sensors are bidirectional. Measurement accuracy is not affected by actual flow
direction or the setting of the
Flow Direction
parameter.
Procedure
Set
Flow Direction
to the value you want to use.
Configure process measurement
40
Micro Motion
®
Model 2700 Transmitters with Analog Outputs
Содержание MICRO MOTION 2700
Страница 10: ...Getting Started 2 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 14: ...Before you begin 6 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 26: ...Configuration and commissioning 18 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 84: ...Configure process measurement 76 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 142: ...Integrate the meter with the control system 134 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 154: ...Set up the Weights Measures application 146 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 156: ...Operations maintenance and troubleshooting 148 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 214: ...Measurement support 206 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 258: ...Troubleshooting 250 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 327: ...Device Tools Trending Figure C 24 Using ProLink III with the transmitter Configuration and Use Manual 319 ...
Страница 328: ...Using ProLink III with the transmitter 320 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 354: ...Default values and ranges 346 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 373: ......