6.3.3
Configure Frequency Output Scaling Method
Display
OFF-LINE MAINT
>
OFF-LINE CONFG
>
IO
>
CH B
>
SET FO
>
FO SCALE
ProLink II
ProLink
>
Configuration
>
Frequency/Discrete Output
>
Frequency
>
Scaling Method
ProLink III
Device Tools
>
Configuration
>
I/O
>
Outputs
>
Frequency Output
Field Communicator
Configure
>
Manual Setup
>
Inputs/Outputs
>
Frequency Output
>
FO Scaling
Overview
Frequency Output Scaling Method
defines the relationship between output pulse and flow
units. Set
Frequency Output Scaling Method
as required by your frequency receiving device.
Procedure
1.
Set
Frequency Output Scaling Method
.
Option
Description
Frequency=Flow
(default)
Frequency calculated from flow rate
Pulses/Unit
A user-specified number of pulses represents one flow unit
Units/Pulse
A pulse represents a user-specified number of flow units
2.
Set additional required parameters.
•
If you set
Frequency Output Scaling Method
to
Frequency=Flow
, set
Rate Factor
and
Frequency Factor
.
•
If you set
Frequency Output Scaling Method
to
Pulses/Unit
, define the number of pulses
that will represent one flow unit.
•
If you set
Frequency Output Scaling Method
to
Units/Pulse
, define the number of units
that each pulse will indicate.
Calculate frequency from flow rate
The
Frequency=Flow
option is used to customize the frequency output for your application
when you do not know appropriate values for
Units/Pulse
or
Pulses/Unit
.
If you specify
Frequency=Flow
, you must provide values for
Rate Factor
and
Frequency Factor
:
Rate Factor
The maximum flow rate that you want the frequency output to report.
Above this rate, the transmitter will report
A110: Frequency Output Saturated
.
Frequency
Factor
A value calculated as follows:
FrequencyFactor =
x N
RateFactor
T
where:
T
Factor to convert selected time base to seconds
Integrate the meter with the control system
Configuration and Use Manual
111
Содержание MICRO MOTION 2700
Страница 10: ...Getting Started 2 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 14: ...Before you begin 6 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 26: ...Configuration and commissioning 18 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 84: ...Configure process measurement 76 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 142: ...Integrate the meter with the control system 134 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 154: ...Set up the Weights Measures application 146 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 156: ...Operations maintenance and troubleshooting 148 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 214: ...Measurement support 206 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 258: ...Troubleshooting 250 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 327: ...Device Tools Trending Figure C 24 Using ProLink III with the transmitter Configuration and Use Manual 319 ...
Страница 328: ...Using ProLink III with the transmitter 320 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 354: ...Default values and ranges 346 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 373: ......