Connection to service port
Figure C-1:
A
C
D
E
B
A. PC
B. Signal converter
C. Service port terminal 7 (RS-485/A)
D. Service port terminal 8 (RS-485/B)
E. Transmitter, with wiring compartment and power supply compartment opened
Note
This figure shows a serial port connection. USB connections are also supported.
4.
Start ProLink III.
5.
Choose
Connect to Physical Device
.
6.
Set
Protocol
to
Service Port
.
Tip
Service port connections use standard connection parameters and a standard address. You do
not need to configure them here.
7.
Set the
PC Port
value to the PC COM port that you are using for this connection.
8.
Click
Connect
.
Need help?
If an error message appears:
•
Switch the leads and try again.
•
Ensure that you have specified the correct COM port.
•
Check the physical connection between the PC and the transmitter.
C.2.3
Make a HART/Bell 202 connection
You can connect directly to the mA terminals on the transmitter, to any point in a local
HART loop, or to any point in a HART multidrop network.
Using ProLink III with the transmitter
300
Micro Motion
®
Model 2700 Transmitters with Analog Outputs
Содержание MICRO MOTION 2700
Страница 10: ...Getting Started 2 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 14: ...Before you begin 6 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 26: ...Configuration and commissioning 18 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 84: ...Configure process measurement 76 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 142: ...Integrate the meter with the control system 134 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 154: ...Set up the Weights Measures application 146 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 156: ...Operations maintenance and troubleshooting 148 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 214: ...Measurement support 206 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 258: ...Troubleshooting 250 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 327: ...Device Tools Trending Figure C 24 Using ProLink III with the transmitter Configuration and Use Manual 319 ...
Страница 328: ...Using ProLink III with the transmitter 320 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 354: ...Default values and ranges 346 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 373: ......