Important
Consult Micro Motion before performing a temperature calibration. Under normal circumstances,
the temperature circuit is stable and should not need an adjustment.
Procedure
See
.
Temperature calibration using ProLink II
Figure 11-13:
Enter temperature of low-
temperature fluid
Temperature Offset calibration
Do Cal
Wait until sensor achieves
thermal equilibrium
Fill sensor with low-
temperature fluid
Calibration in Progress
light turns green
Calibration in Progress
light turns red
Close
Enter temperature of high-
temperature fluid
Temperature Slope calibration
Do Cal
Wait until sensor achieves
thermal equilibrium
Fill sensor with high-
temperature fluid
Calibration in Progress
light turns green
Calibration in Progress
light turns red
Close
Done
ProLink Menu >
Calibration >
Temp offset cal
ProLink Menu >
Calibration >
Temp slope cal
Measurement support
204
Micro Motion
®
Model 2700 Transmitters with Analog Outputs
Содержание MICRO MOTION 2700
Страница 10: ...Getting Started 2 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 14: ...Before you begin 6 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 26: ...Configuration and commissioning 18 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 84: ...Configure process measurement 76 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 142: ...Integrate the meter with the control system 134 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 154: ...Set up the Weights Measures application 146 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 156: ...Operations maintenance and troubleshooting 148 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 214: ...Measurement support 206 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 258: ...Troubleshooting 250 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 327: ...Device Tools Trending Figure C 24 Using ProLink III with the transmitter Configuration and Use Manual 319 ...
Страница 328: ...Using ProLink III with the transmitter 320 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 354: ...Default values and ranges 346 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 373: ......