12.5
Temperature measurement problems
Temperature measurement problems and recommended actions
Table 12-5:
Problem
Possible causes
Recommended actions
Temperature reading
significantly different
from process temper-
ature
•
RTD failure
•
Wiring problem
•
Check junction box for moisture or verdi-
gris.
•
Perform RTD resistance checks and check
for shorts to case (see
).
•
Confirm the temperature calibration factor
matches the value on the sensor tag.
•
Refer to status alarms (especially RTD fail-
ure alarms).
•
Disable external temperature compensa-
tion.
•
Verify temperature calibration.
•
Check the wiring between the sensor and
transmitter. See
.
Temperature reading
slightly different from
process temperature
•
Sensor temperature not yet equalized
•
Sensor leaking heat
•
The RTD has a specification of ±1 °C. If the
error is within this range there is no prob-
lem. If the temperature measurement is
outside the specification for the sensor,
contact Micro Motion.
•
The temperature of the fluid may be
changing rapidly. Allow sufficient time for
the sensor to equalize with the process flu-
id.
•
Insulate the sensor if necessary.
•
Perform RTD resistance checks and check
for shorts to case (see
).
•
The RTD may not be making good contact
with the sensor. The sensor may need to
be replaced.
Troubleshooting
Configuration and Use Manual
223
Содержание MICRO MOTION 2700
Страница 10: ...Getting Started 2 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 14: ...Before you begin 6 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 26: ...Configuration and commissioning 18 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 84: ...Configure process measurement 76 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 142: ...Integrate the meter with the control system 134 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 154: ...Set up the Weights Measures application 146 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 156: ...Operations maintenance and troubleshooting 148 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 214: ...Measurement support 206 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 258: ...Troubleshooting 250 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 327: ...Device Tools Trending Figure C 24 Using ProLink III with the transmitter Configuration and Use Manual 319 ...
Страница 328: ...Using ProLink III with the transmitter 320 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 354: ...Default values and ranges 346 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 373: ......