Page 19 of 80
M8024 Issue 3.4
For use with master product
software V3.5.1.8 and above.
7.8. Communication Protocol Selection (For use with external control).
By default the communication protocol setting is factory set to receive and communicate data via the STEM
protocol. To change the protocol to accept GSIOC ensure your host system is communication platforms
protocol is GSIOC driven.
Follow the following procedure to change the setting.
7.8.1.
With the Integrity 10 connected to mains power and water supply as detailed Section 5, turn on as
detailed in 7.4.
7.8.2.
A splash screen will appear to indicate the unit is functioning and performing a self check activity.
7.8.3.
A home screen will appear. Touch the settings Icon to enter settings mode.
Note:
Setting mode can only be entered immediately after the Integrity has been switch on. Once
run settings have been made the setting icon will disappear.
To en
ter ‘settings mode’ at any time ensure all cells are turned off.
7.8.4.
Touch the communication Icon in the settings screen to enter communication selection.
7.8.5.
From selection screen touch the GSIOC facility selection.