Page 10 of 80
M8024 Issue 3.4
For use with master product
software V3.5.1.8 and above.
5.
INSTALLATION
.
5.1. Electrical safety and installation.
5.1.1.1. This equipment is designed to use safely under the following conditions:-
Indoor use.
Altitude up to 2000 meters.
Temperatures between 5
C and 35
C.
Maximum relative humidity 80% for temperatures up to 25
C decreasing linearly to 50% relative
humidity at 35
C.
Mains supply voltage fluctuations up to
10% of the nominal voltage.
Transient overvoltages typically present on the mains supply. (Overvoltage category II)
Applicable rated pollution degree 2.
The Power Supply Unit (PSU) for the Integrity 10 has been subjected to electrical safety testing as part of
the manufacturing process. However, it is acknowledged that some end users may wish to undertake further
electrical safety testing as part of the installation process.
AC
dielectric withstand (Flash / Hi Pot)
must not
be undertaken on this equipment as damage
may result. Such testing will invalidate warranty.
DC
dielectric withstand (flash / Hi Pot) may be applied to the
PSU only
(i.e. The
Integrity 10 unit must not
be connected when the PSU is being tested. The Integrity 10 unit operates at low voltage (15V DC) and
must not be subjected to dielectric withstand (Flash / Hi Pot) testing.
DC dielectric withstand (flash / Hi Pot) testing shall only be undertaken by competent trained personnel
using suitable test equipment / environment. DC test voltage should be no more than 1.414 times normal
AC test value for a class I (grounded) equipment.
The test equipment / process should incorporate a discharge procedure to remove any stored DC charge
that may remain on the PSU.
5.1.2.
This equipment must be earthed / grounded to a fixed earth / grounded mains socket outlet.
The mains supply is to be earthed / grounded in accordance with current international and local
legislation as applicable in the country / state of operation.
5.1.3.
Ensure only the correct rated mains input fuses are fitted. (Where applicable ensure the correct
Mains cord and moulded IEC plug and lead set fuse if fitted). See Technical Specification Section 8 of
this Instructions book.
5.1.4.
Check the voltage on the product data label on this product unit and those of any accompanying
electrical accessory. Ensure the rating conforms to your local supply.
5.1.5.
It is recommended this product should be connected to a mains supply source which incorporates
an RCD or GFCI device that has a tripping current of 30mA or less. The RCD or GFCI residual Current
Device cuts off power to the equipment immediately it detects a current leakage fault.
5.1.6.
Do not install this product or accessories on a surface which may become wet from
use with other equipment.
5.1.7.
The unit is supplied with a Mains cord and moulded plug wired as follows.
Green / Yellow
or
Green
=
Earth / Ground
Blue
or
White
=
Neutral
Brown
or
Black
=
Live / line hot.