7.2. Reduction Sleeves. Fitting and Removal.
(Reduction sleeves are an accessory available to accommodate 24mm & 16mm diameter glassware
–
see Section 11).
Always ensure the cell temperatures are at a safe to touch. (Between 40° and -5°C).
Always turn off the coolant water supply before removing the drip tray.
Insert reduction sleeve.
7.2.1.
Insert reduction sleeves by pushing them straight into the cells. Reduction sleeves may be fitted into
any position.
Remove reduction sleeve.
7.2.2.
Release the latch on the top of the Integrity unit and lift off the drip tray.
7.2.3.
Push in the oil seal retention clip and lift out oil seal.
7.2.4.
Lift reduction sleeves out of cells.