Page 30 of 80
M8024 Issue 3.4
For use with master product
software V3.5.1.8 and above.
7.11.9.
Target temperature
. The target temperature it the temperature the user wishes the profile to attain.
This can be a minus temperature as well as a positive one. Touch the field to enter the required
setting.
7.11.10. Use the keypad to make the field entry. Press the green tick button to except the set value.
Note:
To enter -25.6 Press 2, then 5, then 6, followed by the minus sign.
7.11.11.
Set Maximum Upper Limit.
To help reduce exothermic reaction when mixing certain chemicals a
maximum upper temperature limit may be set. When the Integrity10 senses an unregistered change in
temperature state action will be taken to prevent the change exceeding the set maximum temperature
limit. Touch the
‘green tick’ to proceed or, touch the ‘red cross’ to commence setting the next profile
entry field. The set default of this field is 165.0°C.