9.4. Cell Replacement.
ATTENTION: If multiple cells are being replaced, replace one cell at a time and follow
the procedure as detailed commencing with the failed cell in the lowest position number. Once
complete replace the next cell and so on.
9.4.1.
Turn off the Integrity. Disconnect it from the Power Supply Unit. Disconnect
the chilled water supply
.
9.4.2.
Release the drip tray by pulling forward on the release catch against the spring pressure.
9.4.3.
Lift the drip tray clear of the Integrity 10 unit by lifting the front end and then disengaging the back of
the drip tray from under the rear handle plate lip.
9.4.4.
Using a Torx screwdriver, carefully unscrew and remove both the front and rear screws on the side
panel adjacent to the cell that is to be replaced.