![Electrothermal INTEGRITY 10 Скачать руководство пользователя страница 67](http://html.mh-extra.com/html/electrothermal/integrity-10/integrity-10_instruction-book_2386095067.webp)
Page 67 of 80
M8024 Issue 3.4
For use with master product
software V3.5.1.8 and above.
9.4.5.
Slide the side panel towards the rear of the Integrity 10 unit between 5 and 10 mm then
Pull
the top edge of the panel away from the unit and then lift side panel clear.
9.4.6.
Using a narrow bladed pair of pliers grip and lift out the two retention pins associated with the faulty
cell. Reserve the pins for re-fitment.
9.4.7.
Lift the cell clear of the Integrity unit assembly and disconnect the IP65 connector plug from the
socket by lifting the latch and pulling apart.