7.10.8.
Time.
The length of time the user wishes the unit to run the cell for is set via this field. By touching
the ‘Time’ field a run time period may be set.
7.10.9. Enter the required run time using the keypad.
Note:
Enter the time in minutes.
Example
, 4 hours 25 minutes is 265 minutes.
Note:
Maximum run time entry is 9999 minutes = 166 hours and 39 minutes.
7.10.10.
Ramp.
The optional field of Ramp rate may be changed using the same procedure. Touch the field
box to select the setting keypad.
Note:
The default setting for ramp rate is 5.0°C per minute. Graduations can be set as low as
0.1°C per minute. When 0.0 is entered, maximum ramp rate is achieved.