![Electrothermal INTEGRITY 10 Скачать руководство пользователя страница 65](http://html.mh-extra.com/html/electrothermal/integrity-10/integrity-10_instruction-book_2386095065.webp)
Page 65 of 80
M8024 Issue 3.4
For use with master product
software V3.5.1.8 and above.
9.2. Spillage Procedure.
In case of solution spillage, turn off the Integrity 10 unit via the Power Supply On / Off switch. For light
spillage wipe off the solution using a water dampened cloth and dry thoroughly before further use.
For spillage involving liquid solution spilt into a reaction well, turn off the Integrity 10 unit via the Power
Supply On / Off switch. Turn off the chilled water supply and disconnect the supply hoses. Uncouple and
remove any peripheral cables linked to a PC or accessories. Remove mains power input and mains
connection cable.
Take unit and tip spillage out of the reaction cells.
(**)
Clean the unit using water dampened cloth and dry
thoroughly.
Ensure the unit is dry before
further use. The Responsible Body shall check the electrical
safety of the unit. Only if all safety requirements are met can the unit be used again. The above
procedure is intended as a guide. Should spillage occur with a toxic or hazardous fluid then special
precautions may be necessary.
(**)
Observe special precautions for handling contaminated substances see
9.3. Decontamination.
If the equipment has been exposed to toxic, radiological or biological contamination, the
Responsible Body is responsible for carrying out appropriate decontamination. If hazardous
material has been spilt on or inside the equipment, decontamination should only be undertaken
under the control of the Responsible Body with due recognition of possible hazards. Before using
any cleaning or decontamination method, the Responsible Body should check with the
manufacturer the proposed method will not damage the equipment.
Decontamination Certificate.
Note:
In the event of any part of this unit becoming damaged, the damaged item(s) should be returned to
the manufacturer for repair accompanied by a decontamination certificate.
Copies of the Certificate are
available from Distributor/Manufacturer. Appendix A of this Instructions book may be copied and
enlarged. At the end of life this product must be accompanied with a Decontamination Certificate.