7.11.23. The default value for number of loops is set at 1. To change the number of times the profile is
repeated, touch the field box and use the keypad to enter the desired value.
Note:
A maximum of 10 loops may be set per profile.
Note:
The start temperature for all subsequent loops will be the same temperature as the finish
temperature of the preceding loop.
7.11.24. By selecting the ramp loop the user is given the facility to enter via the keypad the number of times
they wish to set up a loop. Use the green Tick to except the loop setting or touch the red cross to clear
the entry field and enter another. Touch the back arrow to return to ramp loop selection screen.
7.11.25. It is quite likely the finishing temperature of each loop will be different to the start temperature. The
next screen provides the user with the option to select whether they want the following loops to
commence with the finish temperature of the previous loop or, commence with the temperature set for
the start of the first loop. The
‘Start new loop from previous finish temperature’ is set as the default.
Touch the green tick button once you have made your selection.