Set the Line Voltage
Before powering on the Versatrax 150
™ system, it is important to check that the input voltage settings
are correct - an incorrect voltage setting will damage the system power supply and the winch controller.
When installing the system in a new location always check the line voltage.
•
Monitor:
Universal - no action required.
•
Computer:
Universal - no action required.
•
Interface Box:
Universal - no action required.
•
Power Supply:
Set the line voltage switch to 115/230VAC and change fuse.
o
For 115 VAC use 15 A MDA type fuse.
o
For 230 VAC use 10 A MDA type fuse.
•
Winch:
Jumpers must be set inside the hand-held controller to switch between 115 VAC and
230 VAC. Refer to the winch manual for instructions.
WARNING: Line Voltage Select
– Failing to select the proper line voltage on the
power supply and winch will result in equipment damage.
Generators / Inverters
If powering the system from a generator or inverter, refer to that unit’s operating manual for
recommendations on continuous and peak load ratings. These power sources may apply a reduced
output rating based on electrical load and environmental temperature. Remember to include the power
needs of all other connected devices (external monitors, recording devices, lighting, etc.) when selecting
a generator or inverter.