Measuring principles of sensors
Air temperature sensor
NTC x 2
Skin temperature sensor
NTC
O
2
concentration
galvanic cell
Air humidity
capacitive
Noise level inside the canopy
<49 dB (A) at 50 Hz;
<56 dB (A) at 60 Hz
Air filter
Particle class P 2, in accordance with DIN 3181, factor 2 %
Skin temperature control
Sensor
Use only authentic Dräger sensors (see Order List, p. 57)
Certification marks*
Display component,
15.22
subject to calibration
87.06
Sensor** subject to
calibration
15.14
(2M 20 737)
91.10
Certification document*
Display component,
lB.13-78/87
subject to calibration
Sensor** subject to calibration
IB.9.13-205/91
Calibration tolerances*
Display component,
±
0.2
°
C
subject to calibration
Sensor** subject to calibration
±
0.1
°
C
Total measuring system
±
0.3
°
C
Classification
The incubator complies with DIN-VDE 0750, Part 1,
EN 60601-1, IEC 601-1 and IEC 601-2-19
m
Type B
Electromagnetic compability (EMC)
according to EN 60601-1-2
BabyLink Incubator (optional)
serial interface for output of incubator
data (actual values, set values, alarms)
ThermoMonitoring (optional)
2nd skin temperature sensors to display peripheral skin
temperature.
Dimensions
(width x depth) mm
960 x 680
height in mm
1325 to 1525
Weight
89 kg
* applies only in Germany
** or certified sensor with the same characteristic (400 Series, 2252
W
@ 25
°
C)
53
Technical data