1 Remove mattress from bed.
2 Lift off bed. Undo screw on ventilator hose support
and remove hose support. Remove bed extension.
3 Pull both support brackets up and out.
4 Remove base plate.
Caution. Be careful not to burn yourself.
The risk from the heater is particularly great when the
incubator has been closed and this risk remains for a
long time after the unit has been switched off. (After one
hour heater temperature can still be 70
°
C).
5 Remove fan.
6 Remove trough.
Cleaning/disinfecting/sterilizing
For users in the Federal Republic of Germany we recom-
mend that only disinfectants on the current DGHM
(DGHM: German Society for Hygiene and Microbiology)
list of surface disinfectants are used.
For medical and hygienic reasons, disinfectants based on
quaternary compounds should be used.
Damage may occur to the material of the incubator, parti-
cularly to the sensetive covering parts (PMMA/plexiglass
and PC/polycarbonate), if preparations are used which
are based on
– chlorine-releasing
compounds
– organic and inorganic acids
– alcohols.
With regard to material compatibility we recommend the
following surface disinfectants:
Demykosan AF Messrs. Bayrol, Munich
Incidur
Messrs. Henkel, Düsseldorf
Sekusept Powder Messrs. Henkel, Düsseldorf
Follow manufacturer's instructions.
A precondition for using non-recommended disinfectants
is that the manufacturer can prove material compatibility
with PMMA and PC.
Care
Stripping down
Cleaning/disinfecting/sterilizing
38
1
2
3
3
4
5
6