Wait for warming-up time.
Wait for about 35 minutes before use.
●
Heat incubator in "air temperature control" operating
mode.
If incubator has to be used urgently, operate in standby
mode:
●
Set air temperature to 32
°
C to 36
°
C, see p.20.
●
Oxygen control off, see p.33.
●
Humidity control off, see p.30.
Just before placing the baby in the incubator:
Fill water bottles
●
If required use protective gloves.
1 Open back panel and fold down.
2 Disconnect empty water bottles and refill with 500 mL
clean, distilled or demineralised water (hospital
dispensary). Do not use any additives.
●
Refit filled water bottles and close back panel
again.
●
Then disinfect hands.
Use
– DIN 58363-IL 500-FL-BK1 or
– ISO 8536-1-IL 500-CL-HC1 Form A,
infusion bottles only.
If a water container is used:
Fill water container
●
If required use protective gloves.
1 Take cover of water container off and
2 fill container to mark with clean, distilled or
demineralised water (hospital dispensary).
Do not use any additives. Take care not to overfill.
●
Put cover of water container back.
●
Then disinfect hands.
Placing baby in the incubator
●
Open front door and pull out bed.
●
Put the baby on the bed and slide the bed back in
position.
●
Close the front door and check that the catches are
properly engaged.
●
Adjust the bed as required.
Adjusting working height
●
Adjust working height, as required, see p. 8.
19