Canopy, inside and outside
Inner wall (folded down or removed)
Mattress
Bed and bed extension
Support brackets
Base plate
Trough
Incubator housing, inside and outside
Climate sensor:
●
Remove obvious soiling with disposable cloth and
detergent.
●
Wipe-disinfect surfaces.
●
Allow time for the disinfection process, then wipe the
surfaces again with a clean, damp cloth and dry.
●
Do not allow any liquid to get into climate sensor.
Fan
Seals
Hose seals
Connecting hose
Bottle holder or water container
●
Wash with detergent and rinse with clean water.
●
Disinfect in bath. Allow time for the disinfection
process, then rinse with clear water and dry;
or
●
sterilize at 120
°
C (glove programme).
Use recommended cleaning and disinfecting agents only.
If other agents, such as alcohol for instance, are used
there is a risk of tension cracks in acrylic sheet and
Makrolon.
Do not spray or bath disinfect climate sensor.
Do not use UV-radiation on the incubator. This may cause
cracks in the acrylic sheet.
39
Care
Cleaning/disinfecting/sterilizing