Extending range of set values from 37
°
C to 39
°
C
Only do this if prescribed by a doctor.
The patient's core temperature must be monitored
very carefully.
1 Press unlocking key,
7
,
2 yellow >37
°
C control LED is lit.
The alarm limit for high temperature will be increased
from 38
°
C to 40
°
C.
3 Keep
P
or
p
key pressed until the value required
is displayed.
●
Set a value between 37
°
C and 39
°
C within 1minute.
Otherwise:
– >37
°
C control LED will go out.
The lock against setting a higher range of values will
be in force again.
The alarm limit for high temperature will return to
38
°
C again.
If a value below 37
°
C is set after the range of values has
been extended:
– the extension is automatically cancelled.
If the actual temperature remains above 38
°
C, the
high temperature alarm is triggered.
Recommendation to prevent an high temperature alarm:
●
Set value to 37.1
°
C and wait until incubator has
cooled to 37.9
°
C ,
●
then set value below 37
°
C , as required.
21
Operation
Controlling air temperature
Reset
Control
>37 °C
Check
°C
°C LUFT/AIR
3
3
2
1