Preparing oxygen therapy accessories
Oxygen enrichment of incubator air with oxygen control
1 Screw O
2
connecting hose underneath incubator.
●
Connect probe to terminal of medical gas pipeline
O
2
supply in "parking position".
Oxygen enrichment without O
2
control
If oxygen control module is faulty
●
Follow Instructions for Use of equipment being used.
●
Monitor O
2
concentration.
O
2
flowmeter
●
Attach O
2
flowmeter to rail.
●
Fit hose to connector on O
2
flowmeter and
2 to O
2
connector underneath incubator.
●
Push probe on O
2
connecting hose into the medical
gas terminal unit, only as far as "parking position"
initially.
Oxygen limiter
If an oxygen limiter is going to be used:
●
Unscrew connector on O
2
flowmeter and screw
oxygen limiter on.
●
Connect hose.
Preparation
Preparing oxygen therapy accessories
12
2
1