Skin temperature control module (optional)
Fault
Cause
Remedy
Red
±
0.5
°
C LED flashes.
Skin temperature deviating from
Below set value:
Actual value display flashes.
set value by more than
±
0.5
°
C.
check that sensor is properly fitted
Intermittent tone commences.
to patient.
Above set value:
measure core temperature of patient
and inform doctor responsible for
patient immediately.
Red Sensor LED flashes.
Skin temperature sensor not
Check connection and put right, if
Three dashes flash in centre of
properly connected.
necessary.
actual value display.
Continuous tone commences.
Faulty skin temperature sensor.
Replace sensor.
Err flashes in actual value display.
Fault in skin temperature control
Switch to air temperature control.
Red Inop. LED is lit in main module.
module while in operation.
Call DrägerService.
Continuous tone commences.
Err is lit in actual value display.
Fault in the skin temperature
Other modules continue to operate
control module; skin temperature
properly.
control switched off.
Call DrägerService.
Reference temperature of 36
°
C
Temperature measuring unit not
Call DrägerService.
is outside tolerance of
±
0.1
°
C operating
correctly.
when
6
key is pressed.
Three dashes are lit at the top in
Temperature measured above 42
°
C.
Cooling required.
actual value display.
Inform doctor responsible for
patient immediately.
Three dashes are lit at bottom of
Temperature measured below 30
°
C.
Check that sensor is properly fitted
actual value display.
to patient.
Fault, Cause, Remedy
Skin temperature control module
43
Check
36 °C
±0,1
Control
°C HAUT/SKIN/PEAU/PIEL
± 0,5 °C
Sensor
Control
15.22
87.06
°C
°C