11. Carefully insert the new lamp into the optical bench and tighten the three
screws that secure the lamp to the optical bench. (The replacement lamp is
pre-aligned in the mount.)
12. Squeeze the clip on the Vis lamp connector and insert it into the 4-pin
bulkhead connector on the side wall. Route the wires through the wire guide
in the optical bench (see Figure 5-3).
13. Replace the lamp cover.
14. Thread the cell inlet tubing through the tubing guide on the lamp cover and
reinstall the tubing bracket
15. Replace the flow cell cover and reconnect the inlet and outlet tubing.
16. Replace the detector front panel.
17. Turn on the detector power.
18. Start the PeakNet 6 or Chromeleon server.
19. Reset the lamp lifetime as follows:
a. Press F8 to display the Commands dialog box.
b. Select
UV
(to display UV-related commands).
c. Select
VisLampAge
.
d. Reset the lamp age to zero.
e. Click the
Execute
button.
NOTE Before resuming routine operation, allow 20 to
30 minutes for the lamp to stabilize.
To ensure proper performance, the lamp must be fully seated.
Do not pinch the Vis lamp connector wires under the lamp cover.
Do not pinch the tubing under the flow cell cover.