2 • Description
Doc. 031644-02 9/03
2-3
2.2
Inside Front Panel
NOTE For an illustration of the inside front panel with the flow
cell cover and lamp cover removed, see Figure 5-1.
Grasp the PDA-100 front panel by the sides and pull straight off to access the
inside front panel (see Figure 2-2).
•
The flow cell cover protects the cell from dust, ambient light, fluctuations in
ambient temperature, and vibration.
•
The lamp cover enhances the thermal stabilization of the optical bench. The
optical bench is housed in a compartment behind the inside front panel.
•
Because the flow cell is located behind a cover, leaks from the cell tubing
connections are not immediately apparent. To prevent damage to detector
components, the bottom of the optical bench contains a built-in leak tray. A
sensor in the sump of the leak tray responds to changes in the index of
refraction when wet.
If the leak sensor is activated, it triggers an alarm in the detector and the front
panel
Run
LED begins flashing (see Figure 2-1). In addition, the leak is
reported to the host computer and PeakNet 6 or Chromeleon displays an error
message.
Figure 2-2. PDA-100 Inside Front Panel
Power
Switch
Component Identification
Label
Leak
Sensor
Flow Cell
Cover
Lamp Cover
056345-01
Sensor
Leak
Lamp
Tungsten
Lamp
Deuterium
Cell
Flow
INJURY.
CAUSE PERMANENT EYE
INTENSE LIGHT MAY
INTO LIGHT SOURCES.
AVOID LOOKING DIRECTLY
WARNING
On/Off
Power
Run
Deuterium
Tungsten
Power
Содержание PDA-100
Страница 33: ...3 Operation and Maintenance Doc 031644 02 9 03 3 7 Figure 3 2 PDA 100 3D Control Panel Example ...
Страница 72: ...PDA 100 Photodiode Array Detector 4 16 Doc 031644 02 9 03 ...
Страница 88: ...PDA 100 Photodiode Array Detector A 4 Doc 031644 02 9 03 ...
Страница 98: ...PDA 100 Photodiode Array Detector C 2 Doc 031644 02 9 03 ...