PDA-100 Photodiode Array Detector
4-2
Doc. 031644-02 9/03
4.2
No Detector Response
•
Detector power not turned on
Check that the main power switch is turned on.
Check that the power cord is connected from the rear panel to a power source.
Check the fuses and replace them, if necessary (see Section 5.9).
•
Power-up sequence failed
If the
Power
LED is flashing, turn off the detector power briefly. Turn on the
power again and check that the
Power
LED is lighted, but is not flashing.
•
Lamp not turned on
If a lamp is turned off, the corresponding front panel LED (
Deuterium
or
Tungsten
) is off, also. Turn on the lamp from the software or the TTL input.
•
Lamp LED flashing
The lamp did not power on properly. Press F8 in the software to display the
Commands dialog box, select
UV
(to display UV-related commands), and
check that
UVLampPowerOnSetting
and
VisLampPowerOnSetting
are
Off
.
If the LED continues flashing, replace the lamp(s). See Section 5.7 or
Section 5.8 for replacement instructions for the deuterium lamp or tungsten
lamp, respectively.
•
PDA-100 not detected on the DX-LAN
When the host computer is powered up, the software scans the DX-LAN for
all installed instruments. If the PDA-100 power is off, the detector cannot be
identified. Before turning on the computer, make sure the detector power is
already turned on.
•
DX-LAN interface incorrectly installed
Check the following (for installation instructions, see Section B.3.6):
•
A 10BASE-T DX-LAN cable (P/N 960281) must be connected
between the
DX-LAN
connector on the PDA-100 rear panel and the
Ethernet hub (P/N 056909).
Содержание PDA-100
Страница 33: ...3 Operation and Maintenance Doc 031644 02 9 03 3 7 Figure 3 2 PDA 100 3D Control Panel Example ...
Страница 72: ...PDA 100 Photodiode Array Detector 4 16 Doc 031644 02 9 03 ...
Страница 88: ...PDA 100 Photodiode Array Detector A 4 Doc 031644 02 9 03 ...
Страница 98: ...PDA 100 Photodiode Array Detector C 2 Doc 031644 02 9 03 ...