PDA-100 Photodiode Array Detector
5-12
Doc. 031644-02 9/03
4. When both locks are released, pull the fuse holder straight out of the
compartment and replace the two fuses in the holder with new 4.0 amp fast-
blow IEC 127 fuses (P/N 954763). Dionex recommends always replacing
both fuses.
5. Reinsert the fuse holder into the compartment. (The fuse holder is keyed to fit
only in the correct orientation.) Apply pressure evenly against the holder until
the locks are engaged. When correctly installed, the fuse holder is flush
against the panel.
6. Reconnect the power cord and turn on the power.
Содержание PDA-100
Страница 33: ...3 Operation and Maintenance Doc 031644 02 9 03 3 7 Figure 3 2 PDA 100 3D Control Panel Example ...
Страница 72: ...PDA 100 Photodiode Array Detector 4 16 Doc 031644 02 9 03 ...
Страница 88: ...PDA 100 Photodiode Array Detector A 4 Doc 031644 02 9 03 ...
Страница 98: ...PDA 100 Photodiode Array Detector C 2 Doc 031644 02 9 03 ...