5 • Service
Doc. 031644-02 9/03
5-9
5.8
Replacing the Tungsten Lamp
1. Stop the PeakNet 6 or Chromeleon server.
2. Turn off the PDA-100 power switch.
3. Grasp the detector front panel by the sides and pull straight off to remove.
4. Remove the tubing bracket from the lamp cover (see Figure 5-1).
5. Remove the flow cell inlet tubing from the tubing guide in the lamp cover (see
6. Remove the flow cell cover from the inside front panel (see Figure 5-1).
7. Move the flow cell tubing out of the way. Remove the four screws holding the
lamp cover in place, and then remove the cover and set it aside.
8. Squeeze the clip on the Vis lamp connector and disconnect it from the 4-pin
bulkhead connector on the side wall (see Figure 5-3).
9. Use a screwdriver to loosen the tungsten lamp screws, and then pull the
mounting assembly away from the optical bench.
NOTE The screws are captive in the lamp base and do not need
to be removed.
10. Inspect the new tungsten lamp (P/N 056123T) for fingerprints and dust. If
necessary, clean the lamp with isopropyl alcohol (IPA).
The lamp housing and base may be hot to the touch, especially after
the lamp has been in operation for a long time. Wait until the lamp has
cooled down before continuing.
La lampe et la base de la lampe peuvent être chaudes au toucher,
particulièrement après que la lampe a été allumée pendant longtemps.
Attendez que la lampe ait refroidi avant de continuer.
Das Lampengehäuse und die Lampenbefestigung können sehr heiß
werden, besonders wenn die Lampe längere Zeit in Betrieb war.
Warten Sie, bis die Lampe abgekühlt ist, bevor Sie diese berühren.
Содержание PDA-100
Страница 33: ...3 Operation and Maintenance Doc 031644 02 9 03 3 7 Figure 3 2 PDA 100 3D Control Panel Example ...
Страница 72: ...PDA 100 Photodiode Array Detector 4 16 Doc 031644 02 9 03 ...
Страница 88: ...PDA 100 Photodiode Array Detector A 4 Doc 031644 02 9 03 ...
Страница 98: ...PDA 100 Photodiode Array Detector C 2 Doc 031644 02 9 03 ...