3 • Operation and Maintenance
Doc. 031644-02 9/03
3-21
Example Chromatogram: Selecting the Reference Wavelength
Figure 3-5 shows results obtained from two different channels within the
same run. When the reference wavelength is turned off, baseline drift is -
120
µ
AU/hour; when a reference wavelength (active mode) is selected,
baseline drift is reduced to -40
µ
AU/hour.
The operating conditions are:
Sample wavelength = 520 nm
Sample bandwidth = 10 nm
Reference bandwidth = 50 nm (active reference mode only)
Flowing water at 1.0 mL/min, 10.35 MPa (1500 psi)
Figure 3-5. Effect of Reference Wavelength on Baseline Drift
Reference Wavelength Mode: “Off”
Baseline Drift: -120
µ
AU/hour
mAU
min
min
mAU
Reference Wavelength Mode: “Active”
Baseline Drift: -40
µ
AU/hour