4.15 Diagnostics
4.15.1 Moduleware Run-Time Diagnostics
The PDA-100 Moduleware periodically checks the status of certain
system parameters. Minor problems that do not justify stopping operation
are reported to the host computer and recorded in the Audit Trail. The
Audit Trail includes the date, time, and error code for each problem the
PDA-100 reports. There are two ways to review this information:
•
Check the Audit Trail displayed in the PDA-100 control panel,
or
•
Retrieve the Audit Trail from a sequence by selecting the Audit Trail
folder and the file for that date.
4.15.2 Software Diagnostics and Calibration
PeakNet 6 and Chromeleon contain comprehensive diagnostics for the
PDA-100. This section briefly describes some of these diagnostics; for a
full description, refer to the software Help or user’s manual.
•
Lamp age
Indicates the total number of hours that each lamp has been turned on
in its lifetime.
To reset the lamp age after installing a new lamp:
a. Press F8 to display the Commands dialog box.
b. Select
UV
(to display UV-related commands).
c. Select
UVLampAge
or
VisLampAge
.
d. Reset the number to zero.
e. Click the
Execute
button.
NOTE When the detector power is initially turned on,
the lamp age already indicates some elapsed
time; this is the time required for factory
calibration and test procedures.
Содержание PDA-100
Страница 33: ...3 Operation and Maintenance Doc 031644 02 9 03 3 7 Figure 3 2 PDA 100 3D Control Panel Example ...
Страница 72: ...PDA 100 Photodiode Array Detector 4 16 Doc 031644 02 9 03 ...
Страница 88: ...PDA 100 Photodiode Array Detector A 4 Doc 031644 02 9 03 ...
Страница 98: ...PDA 100 Photodiode Array Detector C 2 Doc 031644 02 9 03 ...