Doc. 031644-02 9/03
2-1
2 • Description
2.1
Front Panel
Figure 2-1 illustrates the front panel of the PDA-100 Photodiode Array Detector.
The
Power
switch provides on/off control of the detector power. LED indicators
identify the status of four functions: the detector power, the deuterium and
tungsten lamps, and the run. A solid lighted LED indicates that the corresponding
function is in progress. A flashing LED indicates that an error occurred; for an
explanation, check the Audit Trail in PeakNet 6 or Chromeleon.
Figure 2-1. PDA-100 Front Panel
LED
LED Status
Function
Run
Lighted
Data is being acquired by the host computer.
Flashing
An alarm was activated; for example, the leak sensor in
the drip tray may have been triggered.
Off
Data is not being acquired by the host computer.
Deuterium
Lighted
The deuterium lamp is on.
Flashing
A command was issued to turn on the deuterium lamp,
but the lamp failed to turn on (see Section 4.5).
Off
The deuterium lamp is off.
Run
Deuterium
Tungsten
Power