PDA-100 Photodiode Array Detector
3-28
Doc. 031644-02 9/03
3.5
Stand-Alone Analog Operation
This section describes limited, stand-alone detector operation when the PDA-100
is not interfaced to a host computer running either PeakNet 6 or Chromeleon
software. In stand-alone operation, TTL inputs can be used to trigger autozero or
turn the deuterium and tungsten lamps on and off. All other detector parameters
remain at their factory default settings and cannot be changed.
Factory Defaults
The PDA-100 outputs four analog signals at all times, even when not interfaced to
a computer. Data is presented at the analog output channels whenever the detector
power is on. Use a chart recorder or other voltage-measuring device to monitor
the channels.
When the PDA-100 power is first turned on, the analog output settings are at the
factory defaults (see the table below).
These factory default settings are also in effect:
•
Rise time: 2.0 seconds.
•
Lamp status: The deuterium and tungsten lamps are on.
Stand-Alone Operation
The table below lists the TTL inputs and the PDA-100 parameters they control
during stand-alone operation.
Channel
Single
Wavelength
Sample
Bandwidth
Reference
Wavelength
Full-Scale
Output
% Offset
1
210
4
Off
1.0 AU
10%
2
254
4
Off
1.0 AU
10%
3
272
4
Off
1.0 AU
10%
4
520
10
Off
1.0 AU
10%
TTL Input
PDA-100 Parameter
TTL 1 in
Autozero
TTL 2 in
Remote inject
TTL 3 in
Deuterium lamp on/off
TTL 4 in
Tungsten lamp on/off
Содержание PDA-100
Страница 33: ...3 Operation and Maintenance Doc 031644 02 9 03 3 7 Figure 3 2 PDA 100 3D Control Panel Example ...
Страница 72: ...PDA 100 Photodiode Array Detector 4 16 Doc 031644 02 9 03 ...
Страница 88: ...PDA 100 Photodiode Array Detector A 4 Doc 031644 02 9 03 ...
Страница 98: ...PDA 100 Photodiode Array Detector C 2 Doc 031644 02 9 03 ...