3 • Operation and Maintenance
Doc. 031644-02 9/03
3-13
The following screen is displayed.
12. Select the sample
Wavelength
,
Reference Wavelength
, and
Reference Bandwidth
.
NOTE The Reference Wavelength plus one-half the
Reference Bandwidth cannot be outside the 3D
Wavelength range. If the values are outside this
range, the Reference Wavelength setting will default
to the longest wavelength possible given the
bandwidth. In the screen shot above, the selected
parameters are acceptable because 750 nm plus 25
nm (one-half of 50 nm) is less than 800 nm.
13. Click
Next
.