B • Installation
Doc. 031644-02 9/03
B-5
B.3.3
Waste Line
Connect the waste line (P/N 054779) to the PDA-100 rear panel (see
Figure B-1). Place the free end of the line in a waste container below the
level of the lab bench on which the PDA-100 is installed.
B.3.4
Detector Output (Analog)
The PDA-100 provides four analog outputs controlled by absorbance
wavelengths 1 through 4 in PeakNet 6 or Chromeleon software. They
provide a 1 volt output maximum. The software controls the full-scale
absorbance output and offset within the 0 to 1 volt range. Connect a
recorder or an integrator to the
ANALOG OUT
connector on the PDA-100
The drain tube must remain below the drain port. If the drain tube is
above the drain port, the drip tray may overflow inside the detector.