B • Installation
Doc. 031644-02 9/03
B-3
B.3.2
Flow Cell
The flow cell is installed in the detector at the factory. Follow the steps
below to complete the cell installation.
1. Grasp the PDA-100 front panel by the sides and pull straight off to
remove.
2. Connect the cell inlet tubing to the union fitting on the heat exchanger
Figure B-1. PDA-100 Rear Panel
Do not touch the cell windows. If you touch a window, clean it with
denatured alcohol and a clean lens tissue.
Do not use the PEEK flow cell with normal phase or chlorinated
solvents; these solvents will damage the cell.
Do not use the stainless steel flow cell with high pH salts; the salts
can cause corrosion, contamination, and metal leaching.
(+)
OUT
TTL
RELAY OUT
2
1
4
3
2
1
2
1
2
1
12
11
10
9
8
7
6
5
4
3
2
1
DX-LAN
ANALOG OUTPUT
CHANNEL 4
CHANNEL 3
CHANNEL 2
CHANNEL 1
TTL IN (+)
TTL
GND
(+)
056338_01
FOR CONTINUED PRO-
TECTION AGAINST
RISK OF FIRE OR
SHOCK, REPLACE
ONLY WITH SAME
TYPE AND RATING
OF FUSE.
WARNING
DUAL FUSE INPUTS.
REMOVE POWER CORD
BEFORE SERVICING.
H3.15A
250V
X X X XX X X X
TUV
MODEL: PDA-100
INPUT: 100-240V
50/60Hz, 1.5A
SERIAL NUMBER
035779-05
Made in U.S.A.
Dionex Corporation, Sunnyvale, California 94086
An alternative license is available, on reasonable terms, to practice any of the foregoing
methods of use which can be praticed without infringing patented apparatus or products.
Apparatus and products manufactured and sold by Dionex Corporation under one or more
of the following U.S. Patents owned or licensed by Dionex: 4,242,097; 4,252,644; 4,314,823;
4,351,909; 4,383,047; 4,455,233; 4,474,664; 4,482,967; 4,512,245; 4,644,807; 4,700,059; 4,751,004;
4,839,788; 4,939,924; 4,999,098; 5,248,426; 5,271,903; 5,296,115; 5,324,752; 5,358,612; 5,457,626;
5,647,967; 5,660,727, and all equivalents in other countries where issued. Purchaser is granted
a paid-up, non-exclusive license to practice with, and for the useful life of, this particular
apparatus or product, the methods of use in any of the foregoing patents and in U.S. Patents
5,126,272; 4,265,634; 4,414,842 and all equivalents in other countries where issued.
Fuse Holder
Main Power
Receptacle
PDA-100
Serial Number
Relay/TTL
Connectors
DX-LAN
Connector
Analog Outputs
Drain
Tube
LAN Identification
Number
057355-01
LAN ID
PDA-100
IC
P/N 057060
Содержание PDA-100
Страница 33: ...3 Operation and Maintenance Doc 031644 02 9 03 3 7 Figure 3 2 PDA 100 3D Control Panel Example ...
Страница 72: ...PDA 100 Photodiode Array Detector 4 16 Doc 031644 02 9 03 ...
Страница 88: ...PDA 100 Photodiode Array Detector A 4 Doc 031644 02 9 03 ...
Страница 98: ...PDA 100 Photodiode Array Detector C 2 Doc 031644 02 9 03 ...