9.3
I-Code 1 (C270)
Open “I-Code 1” panel by choosing “I-Code 1” in the “Radio > Tag Parameters and
Commands” menu.
9.3.1
Parameters tab
Figure 50: I-Code 1 panel (Parameters tab)
Communication
Bit Coding: to define bit rate used for uplink communication.
Number of time slots: to change the number of time slots used during default read.
Default Read
Fields to read: to define whether only tag memory and/or UID should be read during
default read process (unselected read). The 64-bit UID of an I-Code tag is stored in its
first memory block (32 LSBits) and its second memory block (32 MSBits).
The following options are available:
Memory Only: Read memory blocks only. Memory blocks are reported
sequentially.
UID Only: Read the UID from tag’s first and second memory blocks. The UID
will be reported MSB first.
Both memory and UID: Read the UID and memory blocks. The UID (MSB
first) and then memory blocks will be reported sequentially.
First block and Number of blocks: to choose which blocks of tag memory to read
during default read process (unselected read).
Signal level threshold: a tag will be reported only if its signal level is greater or equal
to this threshold during asynchronous read. All tags will be reported if the threshold is
set to 0.
Customization
If Family Code (in hexadecimal) is not null, only tags with the same code will reply.