3. INSTALLATION
Page 39
© 2005 - 2007 DH Instruments, a Fluke Company
3.3.2
INSTALL REFERENCE TURBO VACUUM AND ROUGHING
PUMPS
Attach the turbo pump to the
vacuum tube that extends
down through the bench
using a KF40 flange and a
the KF40 centering ring with
a screen. Orient the pump
so that its lateral KF16
flange is facing towards the
rear of the bench.
Attach turbo pump controller
cable (on turbo pump
controller) to the turbo pump.
Install PFA vent bleed tube
between the adaptor and the
valve attached to the turbo
pump controller bracket.
Install a KF16 tee onto the
turbo pump KF16 fitting.
Orient the tee so that the
common port is connected to
the turbo pump and the legs
are pependicular to the pump
body.
Install a KF16 x KF25
adapator on the right side
(looking from rear) of the
KF16 vacuum tee.
Position vacuum roughing
pump under the bench (the
two scroll vacuum pumps
provided are identical, so
either one may be used here).
Connect turbo pump KF25 connection to roughing pump using KF25 plastic hose.
Содержание ADCS-601-AF
Страница 10: ...ADCS 601 AF OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 12: ...ADCS 601 AF OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page X N NO OT TE ES S...
Страница 114: ...ADCS 601 AF OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 102 N NO OT TE ES S...
Страница 116: ...ADCS 601 AF OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 104 N NO OT TE ES S...
Страница 146: ...ADCS 601 AF OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 134 N NO OT TE ES S...
Страница 168: ...ADCS 601 AF OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 156 N NO OT TE ES S...
Страница 174: ...ADCS 601 AF OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 162 N NO OT TE ES S...