4. OPERATION
Page 93
© 2005 - 2007 DH Instruments, a Fluke Company
Select the DUT with
<Label>
<King 3682>
and
<Output Label>
<Ps>
or
<Pt>
as desired. Press
[Next].
Enter the serial number of the DUT and other identifying information in the
fields provided at the top of the display. Do not change to
<IEEE Address>
Also verify that the range is correct, then press
[Next]
.
Select the
<King 3682 Ps Validation>
or
<King 3682 Pt Validation>
test
from the list and press
[Finish]
. There are no further initialization steps. The
test will complete initialization and start to run. The status bar updates
during initialization to display the current system operation.