3. INSTALLATION
Page 51
© 2005 - 2007 DH Instruments, a Fluke Company
When
<Leading characters to strip>
is displayed, press
[ENT]
to accept the
value of
<0>
.
3.4.2.3
EDIT AND SELECT PG7601 SETUP FILE
On the PG7601 Terminal front panel, select
[SETUP]
,
<3edit>
.
When
<Edit SETUP file:>
is displayed, press
[ENT]
to accept
<#2 >
.
Then select
<7vac>
,
<4COM2>
. Press
[ESC]
and in response to <
Save as
SETUP file:>
press
[ENT]
to accept
<#2 >
. This returns operation to the MAIN
RUN screen.
Select
[SETUP]
,
<1select>
.
When
<Use SETUP file:>
is displayed, enter
<2>
.
<Initializing extern vac sensor>
should be displayed briefly. If a failure
message is displayed, check COM ports (see Section 3.4.2.1)
3.4.3
PG7601
3.4.3.1
POWER UP
In normal ADCS-601-AF operation, the PG7601 is controlled remotely by
ADCS
Tools
software. For power up and initialization, the PG7601’s PG7000 Terminal
keypad and display are used.
Turn the PG7601 power ON by pressing the power ON/OFF switch on the rear
panel of the PG7000 Terminal. Observe the display as the terminal connects
with the PG7601 Platform, tests, initializes and goes to the main run screen.
If
<….Searching…..>
displays for more than 5 seconds, the communications
between the PG7000 and the PG Terminal are failing. Check that the PG7601 to
PG Terminal cable is properly installed.
If PG7601 fails to reach the main run screen, service may be required. Record
the sequence of operations and displays observed and contact a
DHI
Authorized
Service Provider (see Table 26).
3.4.3.2
INITIALIZATION
VERIFY THE PISTON-CYLINDER MODULE, MASS SET AND MASS
LOADING BELL
PG7601 uses stored piston-cylinder and mass set metrological information to
calculate the reference pressures it defines (see the PG7000 Operation and
Maintenance Manual for more detail). For the pressure values to be correct, the
stored metrological information on the piston-cylinder and mass set must be
correct. Before using PG7601 for accurate pressure definition, the validity of the
stored information should be verified. This consists of comparing the piston-
cylinder and mass set information stored in PG7601 to the information in the
current piston-cylinder and mass set calibration reports.
With a new ADCS-601-AF system, the piston-cylinder module and mass set
information have already been loaded into the PG7601.
To verify the PG7601 piston-cylinder and mass set information, use the viewing
capabilities accessed by pressing
[SPECIAL]
,
<1PC/MS>
on the PG7000 Terminal.
Compare the information contained in the PG7601 piston-cylinder and mass set files
to the information on the piston-cylinder and mass set calibration reports.
VERIFY SETUP FILE
Содержание ADCS-601-AF
Страница 10: ...ADCS 601 AF OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 12: ...ADCS 601 AF OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page X N NO OT TE ES S...
Страница 114: ...ADCS 601 AF OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 102 N NO OT TE ES S...
Страница 116: ...ADCS 601 AF OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 104 N NO OT TE ES S...
Страница 146: ...ADCS 601 AF OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 134 N NO OT TE ES S...
Страница 168: ...ADCS 601 AF OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 156 N NO OT TE ES S...
Страница 174: ...ADCS 601 AF OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 162 N NO OT TE ES S...