ADCS-601-AF™ OPERATION AND MAINTENANCE MANUAL
© 2005 - 2007 DH Instruments, a Fluke Company
Page 50
3.4
POWER UP, INITIALIZATION, VERIFICATION
3.4.1
MAIN POWER SWITCHES
Turn ON the
INSTRUMENTS
power switch on the front, top, left of the Control Cabinet.
The (2) PP3 pressure controllers and the RPM4 reference pressure monitor in the
Control Cabinet should initialize and go their main run screens. If they do not power up,
check from the rear of the Control Cabinet that their power cords are properly connected
and/or that the power switch on the power strip inside the Control Cabinet is ON.
Turn ON the
UTILITY PUMP
power switch on the front, top, left of the Control Cabinet. If
the utility pump does not power up, check the power switch on the pump and turn it ON if
necessary.
If
ADCS Tools
software is currently ON, be sure the current measurement mode is
gauge. Turn ON the
REFERENCE PUMPS
power switch on the front, top, left of the
Control Cabinet. The reference turbo and roughing pumps should NOT start up as
ADCS Tools controls a relay on the circuit so that these pumps power up only when
absolute measurement mode is set.
3.4.2
SET OR VERIFY DEVICE COMMUNICATION PORT AND
OTHER SETTINGS
3.4.2.1
COMMUNICATION PORTS
On the RPM4 A116K front panel use
[SPECIAL]
,
<2remote>
,
<1COM1>
to set
the COM1 port.
On the PPC3-200K A116K/BG15K front panel use
[SPECIAL]
,
<2remote>
,
<1COM1>
and
<2COM2>
to set the COM1 and COM2 RS232 ports.
On the PPC3-700K front panel use
[SPECIAL]
,
<2remote>
,
<1COM1>
and
<2COM2>
to set the COM1 and COM2 RS232 ports.
On the PG7601 Terminal front panel, use
[SPECIAL]
,
<5remote>
,
<1COM1>
,
<2COM2>
(after COM2 select
<1settings>
) and
<3COM3>
, <to set the COM1,
COM2 and COM3 RS232 ports.
The correct settings for all RS232 (COM) ports are.
Baud
9
600
Parity
NONE
Length
8
Stop Bit
1
Terminators
<CR><LF> (PG7601 only)
3.4.2.2
PG7601 EXTERNAL VACUUM GAUGE COMMUNICATION
On the PG7601 Terminal front panel, select
[SPECIAL]
,
<5remote>
,
<2COM2>
,
<3vac>
,
<2user>
.
When
<COM2 meas req string>
is displayed, use the
[
→
]
key to scroll through
characters until a
<p>
is displayed. Then press
[ENT]
followed by
[ESC]
.
Choose
<2yes>
in response to the
<Save edits>
prompt.
When
<Reply conv coef>
is displayed, press
[ENT]
to accept the value of
<1.00
Pa/unit>
.
Содержание ADCS-601-AF
Страница 10: ...ADCS 601 AF OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 12: ...ADCS 601 AF OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page X N NO OT TE ES S...
Страница 114: ...ADCS 601 AF OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 102 N NO OT TE ES S...
Страница 116: ...ADCS 601 AF OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 104 N NO OT TE ES S...
Страница 146: ...ADCS 601 AF OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 134 N NO OT TE ES S...
Страница 168: ...ADCS 601 AF OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 156 N NO OT TE ES S...
Страница 174: ...ADCS 601 AF OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 162 N NO OT TE ES S...