4. OPERATION
Page 63
© 2005 - 2007 DH Instruments, a Fluke Company
the requested pressure has been set and stabilized, a
Ready
indication is given (see Section
4.1.1) and the actual value of the set pressure is displayed in the
<Pressure>
field. In most
cases, the final set pressure is not exactly equal to the target pressure (see Section 4.1.4).
4.2.4.1
SETTING ZERO PRESSURE
Setting zero in absolute measurement mode
In absolute measurement mode, entering a target pressure of zero transitions the
active pressure standard to the FPG8601 if it is not already active, controls down
using the FPG’s PPC3 pressure controller to the pressure defined in
[Setup]
,
[Configuration]
,
[Limits]
and then opens the FPG bypass valve (see Figure 29).
This causes the pressure on the FPG8601 and at the ADCS-601-AF
TEST(+)
port,
to be pulled down as low as possible directly by the reference vacuum pump
system. The pressure achieved is indicated in the
ADCS Control Panel
<Pressure>
field.
Ready
is indicated when the FPG8601 stability criterion is
reached.
Setting zero in gauge measurement mode
The result of entering zero as the target pressure in gauge measurement mode
depends upon a
[Maintenance]
,
[FPG Cal Options]
selection (see Section 4.3.2).
If the
<Vent using the FPG in gauge measurement mode>
option IS NOT
selected, the active pressure standard is transitioned to PG7601 if it is not already
active and the PG7601’s PPC3 controller vents to atmosphere setting zero gauge
pressure. The ADCS Control Panel
<Pressure>
field is fixed to
<0.000>
. This is the
default setting.
If the
<Vent using the FPG in gauge measurement mode>
option IS selected, the
active pressure standard is transitioned to FPG8601 if it is not already active and the
FPG8601’s PPC3 controller vents to atmosphere setting zero gauge pressure.
The ADCS Control Panel
<Pressure>
field indicates the reading of the FPG8601.
4.2.4.2
SETTING IDLE STATUS
This section describes the ADCS-601-AF idle condition. The system can be
left in this condition indefinitely but the system must remain powered ON
and supplied with pressure and vacuum. See Section 3.7 for information on
ADCS-601-AF system short and long term shut down procedures.
In the normal idle (resting) condition of the ADCS-601-AF system, the pressure
applied to the
TEST(+)
port is vented. This can be accomplished by pressing the
ADCS Run Screen Toolbar [VENT]
button (see Section 4.3.5.3) or setting a
target pressure of zero in gauge measurement mode (see Section 4.2.4.1).
When vented in absolute mode, the
ADCS Control Panel
<Pressure>
field
indicates
< --- VENT --- >
and the pressure is
Ready.
When vented in gauge
mode, the
<Pressure>
field indicates
<0.000>
and the pressure is
Ready
.
4.2.5
PRESSURE EXERCISING THE DEVICE UNDER TEST
Many devices require pressure exercising prior to validation or calibration. Generally, their
exercising can be performed quickly and without precise knowledge of the endpoint pressures.
ADCS-601-AF supports a function to allow simple, rapid exercising of the DUT. The exercise
function uses the PG7601’s pressure controller alone rather than the pressure standards.
To run an exercise routine, press the
[Exercise Pressure]
button on the
ADCS Run Screen
Toolbar
(see Section 4.3.5.3). Then fill in the form to define the desired exercise routine
(see Figure 30).
Содержание ADCS-601-AF
Страница 10: ...ADCS 601 AF OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 12: ...ADCS 601 AF OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page X N NO OT TE ES S...
Страница 114: ...ADCS 601 AF OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 102 N NO OT TE ES S...
Страница 116: ...ADCS 601 AF OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 104 N NO OT TE ES S...
Страница 146: ...ADCS 601 AF OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 134 N NO OT TE ES S...
Страница 168: ...ADCS 601 AF OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 156 N NO OT TE ES S...
Страница 174: ...ADCS 601 AF OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 162 N NO OT TE ES S...