7. MAINTENANCE OF TRACEABILITY AND RECALIBRATION
Page 155
© 2005 - 2007 DH Instruments, a Fluke Company
To remove the PG7601 reference vacuum CDG assembly and power supply/read
out, proceed as follows:
Prepare the ADCS-601-AF for removal of the CDG:
Vent the ADCS-601-AF system by pressing the
[VENT]
button on the
ADCS
Run Screen
(see Section 4.2.3). Wait for the operation to complete.
Put the system into gauge mode by pressing the
[MODE]
button on the
ADCS Run Screen
. Wait for the operation to complete.
Press
the
[Pause]
button on the
ADCS Main Tool Bar
to suspend remote
communication within the ADCS-601-AF system (see Section 4.3.3).
Press
the
[ESC]
key on the PG7000 Terminal.
Remove the CDG:
Disconnect the CDG isolation valve (valve 3) 1/8 in. pneumatic actuation line
(
V9
) using the quick connector on the valve.
Disconnect the electrical cable on the top of the CDG.
Break the KF40 fitting on the AMH-38 vacuum chamber or the KF25 fitting on
the adaptor. DO NOT BREAK ANY FITTINGS ON THE CDG SIDE OF THE
VALVE.
Remove the CDG supply/readout:
Remove the Control Cabinet main rear panel to access the inside of the cabinet.
Reach inside the cabinet to access the panel mount retainer screws on the
CDG supply/readout. Loosen the screws and remove the supply/readout.
When shipping the CDG for calibration include the CDG, the supply/readout and
the cable that connects the CDG to the supply read/out.
When the CDG and its power supply/readout are received from recalibration,
reinstall them onto the PG7601 and into the ADCS-601-AF control cabinet.
Содержание ADCS-601-AF
Страница 10: ...ADCS 601 AF OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 12: ...ADCS 601 AF OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page X N NO OT TE ES S...
Страница 114: ...ADCS 601 AF OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 102 N NO OT TE ES S...
Страница 116: ...ADCS 601 AF OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 104 N NO OT TE ES S...
Страница 146: ...ADCS 601 AF OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 134 N NO OT TE ES S...
Страница 168: ...ADCS 601 AF OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 156 N NO OT TE ES S...
Страница 174: ...ADCS 601 AF OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 162 N NO OT TE ES S...