ADCS-601-AF™ OPERATION AND MAINTENANCE MANUAL
© 2005 - 2007 DH Instruments, a Fluke Company
Page 94
If the DUT and Test selections are for a
<Ps>
transducer, a prompt asking
whether to determine the
<Offset Check Point>
or not is presented. Press
[Yes]
on the pop-up window to run this step. It is not necessary to run an offset
Check Point when running a test that will determine as received error or when
trying to evaluate the DUT. In this case press
[No]
and skip to Step .
At the next prompt, do NOT press
[OK]
before having set the model 3682 to
<Calibration Manual>
mode. To set
<Calibration Manual >
mode, on the
front panel of the model 3682:
•
Press
[Advance]
•
Press
[Calibration Manual]
•
Press
[Advance]
•
Select
[Calibration]
•
Select
[PS]
or
[PT]
•
Press
[Advance]
•
Select
[Dead Weight Tester]
•
Press
[Advance]
Содержание ADCS-601-AF
Страница 10: ...ADCS 601 AF OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 12: ...ADCS 601 AF OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page X N NO OT TE ES S...
Страница 114: ...ADCS 601 AF OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 102 N NO OT TE ES S...
Страница 116: ...ADCS 601 AF OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 104 N NO OT TE ES S...
Страница 146: ...ADCS 601 AF OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 134 N NO OT TE ES S...
Страница 168: ...ADCS 601 AF OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 156 N NO OT TE ES S...
Страница 174: ...ADCS 601 AF OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 162 N NO OT TE ES S...