5 Operation
Dentsply Sirona
5.1 Acquiring the X-ray image
Operating Instructions Orthophos SL 2D / 3D
5.1.2.4.7
Positioning the patient
The patient is positioned on the unit while standing. Positioning in the seat
is also possible without issue.
Tip: Reference values are shown for the height and forehead support
settings, which are saved for further exposures in the additional
information area of the Sidexis software.
In volume exposures, 2 red laser lines are emitted after switching on the
light localizer. The laser lines show the upper and lower volume limit
depending on the program and collimation.
2 red lines are displayed in the head symbol on the touchscreen. They
indicate the limit of the volume at the bottom and at the top, and the
approximate position of the light beam.
The patient should be aligned in accordance with the occlusal plane. The
position can be easily corrected via the head inclination.
CAUTION
The height adjustment motor starts slowly and then increases its speed.
A patient positioned in the unit may be injured by moving parts.
➢ Monitor the patient and the movement of the unit during height
adjustment. To make minor corrections, press and immediately
release the keys.
CAUTION
The light localizer consists of one Class 1 laser.
Patients and users can be blinded by the laser light localizer.
➢ Do not stare directly into the laser beam. Make sure that the laser
beam does not meet the eyes of the patient.
➢ A distance of at least 10 cm must be maintained between the eye
and the laser.
IMPORTANT
The image quality of volume exposures is limited by metal or other
radiopaque materials located in the patient’s mouth.
Ask the patient to take off all metallic objects such as glasses and
jewelry in the head and neck area as well as all removable dental
prostheses. The tray in front of the control mirror is used for depositing
jewelry.