64 94 954 D3632
D3632
.
201.03.08
.
23 07.2018
103
Dentsply Sirona
5 Operation
Operating Instructions Orthophos SL 2D / 3D
5.1 Acquiring the X-ray image
5.1.2.5.1.3
C3 – Lateral exposure
With this exposure technique, a metal scale integrated in the nose
support is displayed on the X-ray exposure. Using this scale, the
magnification factor in the median plane can be determined precisely via
a measurement.
C3 - Lateral exposure, asymmetric
This program displays a full-format lateral view (approx. 18x23cm). This
program omits the front of the patient's head.
C3F - Lateral full-format exposure
This program displays a full-format lateral view (approx. 30x23cm). This
program displays the whole of the patient's head.
Tip: By default, the image of the lateral exposure C3 or C3F shows the
patient‘s face facing to the right. You can change this representation in
Sidexis.
"Settings"
-
"General Settings"
- Representation -
"Ceph a.p./
p.a."
Please note that all other Ceph exposures C1, C2, and C4 will then also
be displayed ‘mirrored’, i.e. laterally reversed.