5 Operation
Dentsply Sirona
5.1 Acquiring the X-ray image
Operating Instructions Orthophos SL 2D / 3D
5.1.2.4.3
Selecting the volume exposure
✔
The unit is switched on and ready for exposure.
1. Touch the 3D symbol at the top of the touchscreen.
ª
The 3D program group is selected.
2. Select the exposure program. Press the arrow keys + C and - A.
ª
The exposure program is displayed in the program display (B).
The radiation time is displayed under the name of the exposure
program.
3. Follow the instructions in the comment line on the touchscreen. Press
the R key if necessary.
ª
The diaphragm and the sensor move into the starting position for
volume exposures.
ª
The exposure program is selected.
ª
In volume exposures, the total filtration of the X-ray tube assembly is
displayed below the head symbol on the touchscreen. It corresponds
with 0.3 mm Cu in SD and HD modes, and to 1mm Cu in Low mode.
CAUTION
Pressing the R key moves the unit to the starting position.
A patient positioned in the unit may be injured by moving parts.
➢ Check that a patient is not positioned in the unit before moving it to
the starting position.