![Dentsply Sirona Orthophos SL 2D Скачать руководство пользователя страница 62](http://html1.mh-extra.com/html/dentsply-sirona/orthophos-sl-2d/orthophos-sl-2d_operating-instructions-manual_2483684062.webp)
64 94 954 D3632
62
D3632
.
201.03.08
.
23 07.2018
5 Operation
Dentsply Sirona
5.1 Acquiring the X-ray image
Operating Instructions Orthophos SL 2D / 3D
5.1.2.1.8.1
Positioning with an occlusal bite block
The occlusal bite block sets the inclination according to the occlusal plane
instead of the standard Frankfurt horizontal plane. This results in fewer
overlaps in the anterior tooth region and upper jaw.
If required, the service engineer can adjust the angle to the Frankfurt
horizontal plane in order to preset the Frankfurt horizontal plane (see
Service Manual).
✔
The occlusal bite block with bite block foam is inserted in the unit. The
green arrows are displayed on the touchscreen.
✔
The forehead support and temple supports are inserted in the unit.
The hygienic protective sleeves are pulled on.
1. Guide the patient in front of the control mirror.
2. Set the unit height using the up A and down B keys. The height
adjustment buttons are illuminated and show the direction in which
the unit is to be moved by a green arrow.
CAUTION! The height adjustment motor starts slowly and then
increases its speed.
Press and hold down the button until the desired height is reached.
The unit movement is accompanied by a beep.
Release the height adjustment buttons once the bite block plate and
the patient's anterior teeth are at the same height.
3. Guide the patient to the unit and instruct them to hold the handles with
both hands.
4. Instruct the patient to bite into the grooves of the bite block foam with
the teeth.
ª
If necessary, push the lower anterior teeth forwards until they
reach a stop.
B