23 07.2018
5 Operation
Dentsply Sirona
5.1 Acquiring the X-ray image
Operating Instructions Orthophos SL 2D / 3D
5.1.2.5.7.2
Positioning for C3 lateral exposures
✔
The nose support is swiveled upwards.
✔
Push the ear plug holders apart.
✔
The ear plug holders are in a line with the sensor and the secondary
diaphragm.
✔
The protective caps for ear plugs are inserted. The hygienic
protective sleeve for the nose support is pulled on.
1. Set the unit height using the up (A) and down (B) keys.
CAUTION! The height adjustment motor starts slowly and then
increases its speed.
Press and hold down the button until the desired height is reached.
The unit movement is accompanied by a beep.
Release the height-adjustment buttons once the Cephalometer is at
the same height as that of the patient's head.
2. Guide the patient backwards between the ear plug holders.
3. Grasp the ear plug holders at the top and simultaneously slide them
together.
ª
The ear plugs are positioned on the patient's outer auditory
passage.
4. Switch on the light localizer. CAUTION! Risk of dazzle
ª
A red laser line reflects on the patient's head. To switch off the
light localizer, press the key again. It switches off automatically
after approx. 100 seconds.
5. Align the patient's head according to the Frankfurt horizontal plane.
6. Correct the patient's head inclination as necessary. Briefly press the
up (A) and down (B) height adjustment keys.
ª
The laser line reflects on the upper edge of the outer auditory
passage and at the deepest point of the lower eye socket.
B
A