8 Settings and repair
Dentsply Sirona
8.1 Program values
Operating Instructions Orthophos SL 2D / 3D
8.1.3
Volume exposure
The radiation exposure is expressed as an area dose product [mGycm²].
To offset measurement errors and system and unit variations, a tolerance
of 20% must be expected.
The Orthophos SL X-ray system operates at a fixed setting of 85 kV and
values ranging from 4 to 13 mA for volume exposures.
Program:
VOL1 SD
kV/mA
85/7
85/7
85/10
85/13
Effective radiation time
2.6 s
4.4 s
4.4 s
4.4 s
Dose area product (mGycm²) with
whole volume dia. 8 cm x 8 cm
203
338
479
617
Area dose product (mGycm²) with
collimation to dia. 8 cm x 5.5 cm
142
236
334
429
Program:
VOL1 HD
kV/mA
85/4
85/5
85/6
85/7
Effective radiation time
14.4 s 14.4 s 14.4 s 14.4 s
Dose area product (mGycm²) with
whole volume dia. 8 cm x 8 cm
620
781
931
1088
Area dose product (mGycm²) with
collimation to dia. 8 cm x 5.5 cm
431
543
648
757
Program:
VOL1 Low
kV/mA
85/6
85/7
85/10
85/13
Effective radiation time
2.2 s
2.2 s
2.2 s
2.2 s
Dose area product (mGycm²) with
whole volume dia. 8 cm x 8 cm
41
48
67
86
Area dose product (mGycm²) with
collimation to dia. 8 cm x 5.5 cm
28
33
47
60