109
Dentsply Sirona
5 Operation
Operating Instructions Orthophos SL 2D / 3D
5.1 Acquiring the X-ray image
5.1.2.5.6
Setting the kV/mA values
Setting the kV/mA values via the patient symbols
Preset kV/mA combinations are assigned to the patient symbols, which
are selected according to the patient's size and weight. The symbols
roughly correspond to child, youth/woman, woman/man, hefty persons.
✔
Level 1 is displayed on the touchscreen.
➢ Touch the required patient symbol.
ª
The selection (A) is highlighted in orange. The selected kV/mA
value is displayed on the right side of the touchscreen.
ª
The kV/mA value is set.
For the USA only: The source to image distance (SID) is displayed on the
touchscreen. It equals 1.714 mm or 67.5".
Setting the kV/mA values via the submenu line
If the preset kV/mA combinations do not provide a satisfactory result, you
can also set the kV/mA values manually in all programs.
✔
Level 1 is displayed on the touchscreen.
1. Touch the kV/mA symbol (B) on the right side of the touchscreen.
ª
A submenu line is opened.
2. Select a kV/mA value. Touch the - or + keys.
ª
The selected kV/mA value is displayed.
3. Touch the cross on the left side of the submenu line.
or
➢
Touch the kV/mA symbol (B) again.
ª
The submenu line is closed.
ª
The kV/mA value is set.
Ready for exposure
A
B
Ready for exposure
Ready for exposure
Ready for exposure